BEKA BA307C 31/2 digital loop powered 4/20mA indicatorCategory: Indicator/Recorder Manufacturer: Beka Seller's Description: Intrinsically Safe ATEX II 1G EEx ia IIC T5 Operating Temp. Range -20°C to +60°C IP65 The BA307C instrument is loop powered but only induces a 1V drop which allows it to be installed in series with almost any 4-20mA loop. The main application of the BA307C is to display a measured variable or control signal. The zero and span of the display are independantly adjustable so that the indicator can be calibrated to display any linear variable represented by a 4-20mA signal, such as temperature, pressure, level or actuator position. For flow applications an optional square root extractor enables the output from a differential flow transmitter to be displayed in linear engineering units. Back lighting is available as an option to improve readability in poorly illuminated areas. Power Supply Typically via 28V, 300 ohms, Zenner Barrier of Galvanic Isolator Display +/- 1999, 12.7mm high 3 1/2 digits, Backlight (option) Options Backlight, root extractor, calibrator Electrical Connection Plug-in screw terminals for cable 0.5-1.5mm sq Dimensions 48mm (h) x 96mm (w) x 84mm (d) Mounting Front of panel in cut out 45mm (h) + 0.6mm x 90mm (w) + 0.8mm Weight 0.3Kg Guarantee 3 Year Ambient Temperature -20°C to +60°C IP Rating IP65 Front, IP20 Rear Certification Type II 1G EEx ia IIC T5 Certification Number ITS02ATEX2027